| Model Number |
ZKB00 |
| Medical Device Problem Codes |
Bent (1059); Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
03/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lens was not implanted and therefore not explanted.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that upon opening the lens case, the intraocular lens was observed not being in the correct place.Reportedly, the lens was bent or badly scratched.The lens was not used.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation? yes.Returned to manufacturer on: 04/09/2018.Device returned to manufacturer? yes.Device evaluation: the intraocular lens (iol) was returned at the manufacturing site for evaluation.Visual inspection at 12x microscope magnification showed that the lens was contaminated, dust particles were present.This is expected as the lens was transported from controlled environment during manufacturing to a non-sterile, non-environmental controlled area.No other anomalies were found.The customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no similar complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|