• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - CARTAGO INTERLINK SOLUTION ADMINISTRATION SETS; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - CARTAGO INTERLINK SOLUTION ADMINISTRATION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C7557
Medical Device Problem Code Cut In Material (2454)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device was received and evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the blue slide clamp of a vented paclitaxel set severed the tubing which resulted in a leak.The slide clamp severed the tubing during patient setup.The line had been primed and the patient was connected for therapy when the severed tubing was identified.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
The device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional, pressure and clear passage under water testing were performed and the device was found to be within the product specification range.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLINK SOLUTION ADMINISTRATION SETS
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE - CARTAGO
cartago
MDR Report Key7432504
Report Number1416980-2018-02158
Device Sequence Number16478640
Product Code FPA
UDI-Device Identifier00085412036373
UDI-Public(01)00085412036373
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K160007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 05/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/23/2021
Device Catalogue Number2C7557
Device Lot NumberR16H23098
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2018
Initial Date Received by Manufacturer 03/22/2018
Supplement Date Received by Manufacturer05/08/2018
Initial Report FDA Received Date04/16/2018
Supplement Report FDA Received Date05/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DEXTROSE
Patient Age62 YR
-
-