PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. ICT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number 728306
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.(b)(4).
 
Event Description
This complaint has been evaluated based on the information provided; there is no allegation of death or serious injury.The customer reported the patient voice cannot be heard.If the operator is unable to hear the patient due to a failed component, there is potential for injury to the patient.This issue has been determined to be a reportable event.This event is currently under investigation.
 
Manufacturer Narrative
The customer reported that the patient voice cannot be heard.This occurred on the customers brilliance ict system.There was no report of patient harm as a result of this reported issue.The philips field service engineer (fse) arrived on site and determined the breathing light assembly had failed and needed to be replaced to correct the reported issue.The philips fse replaced the breathing light assembly part and the system returned to full functionality.The system is in clinical use and operational.This event has been determined not to be a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICT
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 miner rd
cleveland OH 44143
MDR Report Key7432715
MDR Text Key105695117
Report Number1525965-2018-00390
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
PMA/PMN Number
K060937
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/16/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number728306
Device Catalogue NumberNCTC780
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/20/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-