• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. TRUETRACK; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. TRUETRACK; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUETRACK
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(manufacturer narrative = t, corrected data = f) (b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-20-user's test strip had poor storage (kitchen).Test strip (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
 
Event or Problem Description
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 483 and 373 mg/dl.The customer's expected fasting blood glucose test result range is 170 - 180 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer.The product is not stored according to specification and is stored in the kitchen.The test strip lot manufacturer's expiration date is 08/16/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (b)(6).
 
Event or Problem Description
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 483 and 373 mg/dl.The customer's expected fasting blood glucose test result range is 170 - 180 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer.The product is not stored according to specification and is stored in the kitchen.The test strip lot manufacturer's expiration date is (b)(6) 2019 and open vial date is february 2018.The meter memory was reviewed for previous test result history: result 1 :169mg/dl date: (b)(6) 2018 time:08:22am fasting.Result 2 :151mg/dl date: (b)(6) 2018 time:05:12am fasting.Result 3 :200mg/dl date: (b)(6) 2018 time:08:20am fasting.Result 4 :373mg/dl date: (b)(6) 2018 time:08:20am fasting.Result 5 :483mg/dl date: (b)(6) 2018 time:08:07am fasting.
 
Additional Manufacturer Narrative
(manufacturer narrative = t, corrected data = f) internal report # (b)(4) returned meter evaluated with no defect found.Returned test strips evaluated with defect found.Qc investigation on 5/3/2018 resulted in defect found; s button missing on returned meter.Reported defect not reproduced with returned meter and test strip.Final root cause : rc-003 other root cause: s button cover missing due to user/mechanical stress test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUETRACK
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7433369
Report Number1000113657-2018-00433
Device Sequence Number1544848
Product Code NBW
UDI-Device Identifier00021292001018
UDI-Public(01)00021292001018
Combination Product (Y/N)N
Initial Reporter StateCA
PMA/510(K) Number
K032657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 05/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/16/2019
Device Model NumberTRUETRACK
Device Catalogue NumberA4H01-81
Device Lot NumberRU5091
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/02/2018
Was the Report Sent to FDA? No
Event Location Home
Initial Date Received by Manufacturer 03/22/2018
Supplement Date Received by Manufacturer03/22/2018
Initial Report FDA Received Date04/16/2018
Supplement Report FDA Received Date05/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SECOND THERAPY; SECOND THERAPY
-
-