• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEODENT - JJGC S.A. CM TITAMAX EX ACQUA IMPL TI 4.0X13; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEODENT - JJGC S.A. CM TITAMAX EX ACQUA IMPL TI 4.0X13; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 140.638
Medical Device Problem Codes Failure to Osseointegrate (1863); Loss of Osseointegration (2408)
Health Effect - Clinical Code Tissue Damage (2104)
Date of Event 03/27/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Exemption number: e2015015.
 
Event or Problem Description
The clinician reported that 7 months after the dental implant and abutment were placed in ada site 7, the implant had not achieved osseointegration in type ii bone.Clinician reports deficient oral hygiene.The product will be forwarded to the manufacturer for investigation.There were no reported complications.
 
Additional Manufacturer Narrative
Exemption number: e2015015.Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.In addition, it was seen that the dentist has used more drills than recommended to perform the implant installation.
 
Event or Problem Description
The clinician reported that 7 months after the dental implant and abutment were placed in ada site 7, the implant had not achieved osseointegration in type ii bone.Clinician reports deficient oral hygiene.The product will be forwarded to the manufacturer for investigation.There were no reported complications.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CM TITAMAX EX ACQUA IMPL TI 4.0X13
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
NEODENT - JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key7434543
Report Number3008261720-2018-01772
Device Sequence Number18871962
Product Code DZE
UDI-Device Identifier07899878016571
UDI-Public07899878016571
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K133592
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Dental Assistant
Type of Report Initial,Followup
Report Date (Section B) 04/16/2018
Report Date (Section F) 04/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Dentist
Device Expiration Date01/31/2019
Device Catalogue Number140.638
Device Lot Number91090
Other Device ID Number07899878016571
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2018
Was the Report Sent to FDA? No
Distributor Facility Aware Date06/26/2018
Device Age1 MO
Event Location Other
Date Report to Manufacturer04/02/2018
Initial Date Received by Manufacturer 04/02/2018
Supplement Date Received by Manufacturer06/11/2018
Initial Report FDA Received Date04/16/2018
Supplement Report FDA Received Date06/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ABUTMENT REMOVAL; ABUTMENT REMOVAL; ABUTMENT REMOVAL
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
-
-