| Model Number |
AED PLUS |
| Medical Device Problem Codes |
No Display/Image (1183); Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device would not power up.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental medwatch report is reporting the evaluation of the device.This supplemental medwatch report is also correcting information submitted on the initial medwatch report.Device evaluation: the malfunction was observed during incoming testing; however after disassembling the device to determine root cause, the malfunction was no longer seen.The device was put through extensive testing without duplicating the malfunction.The led interface assembly board was replaced as a precaution.The device passed the final test procedure, was recertified, and returned to the customer.Analysis for reports of this type has not identified an increase in trend.Review of activity logs did not find evidence to support the reported event.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device powered on without display.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|