| Catalog Number |
816085100 |
| Health Effect - Clinical Code |
Overdose (1988)
|
| Date of Event |
03/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This is our initial report on this incident.
|
| |
|
Event or Problem Description
|
|
The customer reported discrepant results between two tango infinity analyzers with biotest cell 3.On one analyzer the result was positive but negative on the other analyzer.The customer stated that four samples of two different patients were affected.Both patients had received rh(d) prophylaxis and due to this samples were supposed to yield positive reactions with the rh(d) positive screening cells #1 and #2 of biotest cell-3.The two samples on each patient were pre- and post-delivery.The customer stated that the positive reactions were +/- respectively 1+ positive.The customer returned the four samples of the two patients (each pre- and post-delivery) that had caused false negatives for investigational testing, but not the supposedly defective product biotest cell 3.Therefore our quality control laboratory tested the patient samples with their retention sample of biotest cell 3 on tango infinity.One pre-delivery sample yielded a correctly weak positive result, but the other pre-delivery sample and both post-delivery sample yielded negative results.Three of the four samples were additionally tested in the gel technique with enzyme treated reagent red blood cells and yielded very weak positive results.Further tests and testing of the fourth patient sample was not possible, because the material was used up.As a summary of all test results the anti-d of the two patients are very weak reacting antibodies.The instruction for use contains a note in the section limitation: "negative reactions will be obtained if the sample contains antibodies present in concentrations too low to be detected by the test method employed.No test method is capable of detecting all red cell antibodies." the retention sample of the supposedly defective lot of biotest cell 3 was tested with different samples and controls on tango infinity and reacted as expected.Testing by our quality control laboratory confirmed that the allegedly defective lot of biotest cell 3 functions correctly.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of both allegedly defective lots.Investigation of the affected tango infinity is still ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
This is our final report on this incident.
|
| |
|
Event or Problem Description
|
|
The customer reported discrepant results between two tango infinity analyzers with biotestcell 3.On one analyzer the result was positive but negative on the other analyzer.The customer stated that four samples of two different patients were affected.Both patients had received rh(d) prophylaxis and due to this samples were supposed to yield positive reactions with the rh(d) positive screening cells #1 and #2 of biotestcell-3.The two samples on each patient were pre- and post-delivery.The customer stated that the positive reactions were +/- respectively 1+ positive.The customer returned the four samples of the two patients (each pre- and post-delivery) that had caused false negatives for investigatioal testing, but not the supposedly defective product biotestcell 3.Therefore our quality control laboratory tested the patient samples with their retention sample of biotestcell 3 on tango infinity.One pre-delivery sample yielded a correctly weak positive result, but the other pre-delivery sample and both post-delivery sample yielded negative results.Three of the four samples were additionally tested in the gel technique with enzyme treated reagent red blood cells and yielded very weak positive results.Further tests and testing of the fourth patient sample was not possible, because the material was used up.As a summary of all test results the anti-d of the two patients are very weak reacting antibodies.The instruction for use contains a note in the section limitation: "negative reactions will be obtained if the sample contains antibodies present in concentrations too low to be detected by the test method employed.No test method is capable of detecting all red cell antibodies." the retention sample of the supposedly defective lot of biotestcell 3 was tested with different samples and controls on tango infinity and reacted as expected.Testing by our quality control laboratory confirmed that the allegedly defective lot of biotestcell 3 functions correctly.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of both allegedly defective lots.The affected tango infinity was checked by our field service engineers with no indication for a malfunction.The instrument is confirmed to run within specifications.
|
| |
|
Search Alerts/Recalls
|