The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure to a defective processor board.The technician replaced the board to resolve the reported issue.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Evaluated on site by livanova rep.
|