| Model Number |
LF1744 |
| Medical Device Problem Codes |
Self-Activation or Keying (1557); Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
03/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during the procedure they connected device to a generator, however an error sound was heard immediately.The generator was replaced by another generator, and the same event occurred.After that, the error sound stopped and the physician used the device as usual, however the sealing was performed without activation.Also, an activation sound was heard without pressing the button when they set the device on the surgical table.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one device was received for evaluation.This device had been used in the treatment or diagnosis of a patient.A review of the lot number reported indicates that the product was within the assigned expiration date at the time of the reported incident.The returned product met specification as received.Visual inspection found no defects.The device was activated ten times on a saline soaked towel with satisfactory results.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.The device was activated with the rotation knob in various positions to detect any problematic activation issues.The alleged incident could not be duplicated.The investigation found the device to function normally and within specifications.The ifu states, keep the instrument jaws clean.Build up of eschar may reduce sealing and/or cutting effectiveness.Wipe jaw surfaces and edges with a wet gauze pad as needed.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|