MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number LF1744
Medical Device Problem Codes Self-Activation or Keying (1557); Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Information (3190)
Date of Event 03/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during the procedure they connected device to a generator, however an error sound was heard immediately.The generator was replaced by another generator, and the same event occurred.After that, the error sound stopped and the physician used the device as usual, however the sealing was performed without activation.Also, an activation sound was heard without pressing the button when they set the device on the surgical table.
 
Additional Manufacturer Narrative
Evaluation summary: one device was received for evaluation.This device had been used in the treatment or diagnosis of a patient.A review of the lot number reported indicates that the product was within the assigned expiration date at the time of the reported incident.The returned product met specification as received.Visual inspection found no defects.The device was activated ten times on a saline soaked towel with satisfactory results.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.The device was activated with the rotation knob in various positions to detect any problematic activation issues.The alleged incident could not be duplicated.The investigation found the device to function normally and within specifications.The ifu states, keep the instrument jaws clean.Build up of eschar may reduce sealing and/or cutting effectiveness.Wipe jaw surfaces and edges with a wet gauze pad as needed.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key7436170
Report Number1717344-2018-00497
Device Sequence Number19186609
Product Code GEI
UDI-Device Identifier20884521181561
UDI-Public20884521181561
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133338
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/28/2022
Device Model NumberLF1744
Device Catalogue NumberLF1744
Device Lot Number72330167X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2018
Initial Date Received by Manufacturer 03/29/2018
Supplement Date Received by Manufacturer05/25/2018
Initial Report FDA Received Date04/17/2018
Supplement Report FDA Received Date06/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-