| Model Number |
SM-644 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Device Packaging Compromised (2916)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a procedure, the needle was not connected to the thread when they opened the sterile package.Then after the operation, they opened several more that was in the same shape.They used the device as soon as they opened it up during the operation.The was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned products noted one suture and one needle.The needle was observed to be detached from the suture.Upon microscopic inspection of the needle, normal needle crimp and swage marks were observed on the needle.The retainer was inspected with no abnormalities.As no sealed products were returned, functional testing was precluded.A review of the device history record indicates the product was released meeting all quality release specifications at the time of manufacture.Unfortunately, because no sealed samples were received, a needle attachment test could not be performed.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|