MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD BIOSYN; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD BIOSYN; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID Back to Search Results
Model Number SM-644
Medical Device Problem Codes Detachment of Device or Device Component (2907); Device Packaging Compromised (2916)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a procedure, the needle was not connected to the thread when they opened the sterile package.Then after the operation, they opened several more that was in the same shape.They used the device as soon as they opened it up during the operation.The was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned products noted one suture and one needle.The needle was observed to be detached from the suture.Upon microscopic inspection of the needle, normal needle crimp and swage marks were observed on the needle.The retainer was inspected with no abnormalities.As no sealed products were returned, functional testing was precluded.A review of the device history record indicates the product was released meeting all quality release specifications at the time of manufacture.Unfortunately, because no sealed samples were received, a needle attachment test could not be performed.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOSYN
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7436327
Report Number9612501-2018-00753
Device Sequence Number1055473
Product Code GAM
UDI-Device Identifier20884521033198
UDI-Public20884521033198
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K000037
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 06/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Model NumberSM-644
Device Catalogue NumberSM-644
Device Lot NumberD7E0321X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2018
Initial Date Received by Manufacturer 03/29/2018
Supplement Date Received by Manufacturer05/23/2018
Initial Report FDA Received Date04/17/2018
Supplement Report FDA Received Date06/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-