Model Number VLOCL0344
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/29/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during laparoscopic inguinal hernia, the thread of the device was broken.They then used another suture with the same code from the same box to resolve the issue and complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Post market vigilance (pmv) led an evaluation of two devices.The visual inspection of the open sample noted one partial suture and one needle.The product was received without the loop.It was observed that the suture was broken.The suture was noted to be attached to a needle.The visual inspection and functional evaluation of the sealed sample had acceptable results.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-LOC 180
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key7436569
Report Number1219930-2018-02045
Device Sequence Number16191976
Product Code GAM
UDI-Device Identifier10884521131033
UDI-Public10884521131033
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K091087
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2020
Device Model NumberVLOCL0344
Device Catalogue NumberVLOCL0344
Device Lot NumberA7M0811VX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2018
Initial Date Received by Manufacturer 03/29/2018
Supplement Date Received by Manufacturer06/01/2018
Initial Report FDA Received Date04/17/2018
Supplement Report FDA Received Date06/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-