• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD VERSAPORT PLUS; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD VERSAPORT PLUS; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 179078P
Medical Device Problem Codes Break (1069); Detachment Of Device Component (1104)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during laparoscopic coelioscopy procedure, after the introduction of the trocar, she realized that the trocar broke a triangular tip was missing.The broken end was found in the patient's abdomen.To complete the case, they removed the tip from the abdomen and replaced the trocar with a other device.No patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of two photographs of a device.The first photograph depicts the device being help during a surgical procedure.The second photograph depicts the tip of the disengaged device being retrieved during a surgical procedure.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of the disengaged trocar tip may occur if the device is mishandled or handled roughly during application.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened, and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAPORT PLUS
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7436621
Report Number9612501-2018-00754
Device Sequence Number1007135
Product Code GCJ
UDI-Device Identifier10884521082663
UDI-Public10884521082663
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K954108
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 05/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number179078P
Device Catalogue Number179078P
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2018
Initial Date Received by Manufacturer 03/29/2018
Supplement Date Received by Manufacturer04/25/2018
Initial Report FDA Received Date04/17/2018
Supplement Report FDA Received Date05/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-