• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN POLYSORB; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN POLYSORB; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID Back to Search Results
Model Number GL-129
Medical Device Problem Codes Break (1069); Product Quality Problem (1506)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, intra-operatively, a ¿surjet¿ suturing technique was employed by the surgeon, but at the end of the open procedure, the first and second suture had thread broken but the other one had needle separated from the thread and fell into the patient cavity and was retrieved by the surgeon.The first doctor took another suture but the needle quickly twisted and then the second doctor took another suture and it worked.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary post market vigilance (pmv) led an evaluation of two devices.One needle was returned disengaged from the suture.Upon microscopic inspection, proper crimp marks were noted at the swaged end.The visual inspection and functional evaluation of the sealed sample had acceptable results.A review of the device history record indicates the product was released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYSORB
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven 06473
MDR Report Key7436749
Report Number1219930-2018-02059
Device Sequence Number1678051
Product Code GAM
UDI-Device Identifier10884521045033
UDI-Public10884521045033
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K963253
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2019
Device Model NumberGL-129
Device Catalogue NumberGL-129
Device Lot NumberA4J1176X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2018
Initial Date Received by Manufacturer 03/30/2018
Supplement Date Received by Manufacturer06/27/2018
08/08/2018
Initial Report FDA Received Date04/17/2018
Supplement Report FDA Received Date07/16/2018
10/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-