| Model Number |
EGIA30AV |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a thoraco/lobectomy procedure, the surgeon approached a blood vessel with the device once, but removed it without firing due to insufficient dissection of the vessel.The device was left in the clean field for later firing.After that, the reload was loaded onto the handle without problem, but when surgeon tried to fire the device to the blood vessel, it could not fire.The status of the patient was fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The reload was pre-fired and engaged in interlock.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Replication of the pre-fire condition with interlock engagement may occur if the instrument firing handle had been partially compressed and released after pressing the green firing button.In this situation, the safety interlock feature will engage and prevent the reload from firing a second time by ceasing the placement of staples and tissue transection, and prevent patient harm.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|