• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL HEAL COLLAR 5.7X6.5, 3MM; HEALING COLLAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL HEAL COLLAR 5.7X6.5, 3MM; HEALING COLLAR Back to Search Results
Catalog Number HC563
Device Problem Positioning Problem (3009)
Patient Problem No Information (3190)
Event Date 03/13/2018
Event Type  malfunction  
Event Description
It was reported that the doctor was unable to seat the healing collar (hc563).
 
Manufacturer Narrative
Zimmer biomet complaint (b)(4).One zimmer dental heal collar was returned for inspection.A visual inspection revealed wear about the collar and the threads are noted to be damaged at the engaging surface.The hex head was also slightly worn, but functional.A device history review was performed and no related nonconformance¿s were noted.A complaint history search was performed using our complaint handling system and there were no additional related complaints against this lot.Appropriate documentation was reviewed and the following information was identified: instructions for use for healing collars, healing screws, surgical cover screws and temporary gingival cuffs 9658 rev 1-01/15.Information identified: healing collars ¿ for use with tapered screw-vent, screw-vent and trabecular metal implants select the appropriate size healing collar for the platform of the implant and with appropriate height and emergence profile to fit the existing or desired tissue contour of the site.The healing collars are anodized with color-coding on the lower portion to indicate the implant platform diameter.The top surface of the healing collar is etched with three numbers to reference implant platform diameter, emergence profile diameter and cuff height.The numbers for implant platform and emergence profile show only the initial digit of the related measurement.The product was functionally tested, and it was confirmed that the collar would not screw into a mating implant.It is clear from the package drawing under pd-hc34563 rev a that the first two threads are compressed upward.The problem with screwing in the collar appears to be due to this damage, as the orientation of the engaging threads does not match the drawing and are deformed due to this damage.The complaint is confirmed.A singular cause cannot be determined.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEAL COLLAR 5.7X6.5, 3MM
Type of Device
HEALING COLLAR
Manufacturer (Section D)
ZIMMER DENTAL
6221 el camino real
carlsbad CA 92009
Manufacturer (Section G)
ZIMMER DENTAL
6221 el camino real
carlsbad CA 92009
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key7437117
MDR Text Key105889846
Report Number0001038806-2018-00419
Device Sequence Number1
Product Code NHA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK111852
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2022
Device Catalogue NumberHC563
Device Lot Number2017111345
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/27/2018
Initial Date FDA Received04/17/2018
Supplement Dates Manufacturer Received08/07/2018
Supplement Dates FDA Received08/22/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/18/2017
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age50 YR
-
-