It was reported that during a dialysis access management procedure on an unspecified arm, the performer introducer valve was found to be leaking after the balloon and wire were removed.A new introducer was inserted and the procedure was successfully completed.According to the initial reporter, the patient did not require any additional procedures.
|
|
Investigation ¿ evaluation: a review of the complaint history, device history record, drawing, instructions for use (ifu), manufacturing instructions, quality control, and a visual inspection and functional evaluation of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed the presence of a slow leak, which was observed to be coming from the silicone disc.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.Measures are being conducted to address this failure mode.We will continue to monitor for similar complaints.Per the quality engineering risk assessment, no further action is required.
|