Catalog Number RBY4C1650
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the pet lock was intact on the proximal end of the pusher assembly.The pusher assembly was fractured approximately 63.5 cm and kinked approximately 5.0 cm from the proximal end.The embolization coil was detached from its pusher assembly.Conclusions: evaluation of the returned device revealed that the ruby coil pusher assembly was fractured.If the device is forcefully advanced against resistance, damage such as a kink may occur.Subsequently, if the kink is then attempted to be straightened, damage such as a fracture may occur.The root cause of the inability to advance the ruby coil through the lantern could not be determined.The lantern referred to in the complaint was not returned to penumbra for evaluation.Penumbra catheters and coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using ruby coils.During the procedure, the physician manually flushed the lantern delivery microcatheter (lantern), and then attempted to place a ruby coil.However, the physician was unable to advance the ruby coil past the middle of the lantern.The ruby coil and lantern were therefore removed, and the procedure was then completed using the same lantern and additional ruby coils.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7437392
Report Number3005168196-2018-00794
Device Sequence Number1031256
Product Code HCG
UDI-Device Identifier00814548013244
UDI-Public00814548013244
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date12/07/2023
Device Catalogue NumberRBY4C1650
Device Lot NumberF66862
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2018
Initial Report FDA Received Date04/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age53 YR
Patient Weight86
-
-