Model Number 728306 |
Device Problem
No Audible Alarm (1019)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
We have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.Internal cross reference: complaint (b)(4).
|
|
Event Description
|
This complaint has been evaluated based on the information provided; there is no allegation of death or serious injury.The customer reported an issue with the ct gantry microphone and therefore, the patient cannot be heard by the operator in the control room.If the operator is unable to hear the patient due to a failed microphone, there is potential for injury to the patient.This issue has been determined to be a reportable event.This event is currently under investigation.
|
|
Manufacturer Narrative
|
The customer reported on (b)(6) 2018 that the operator could not hear sounds from the scan room.The philips field service engineer (fse) confirmed there was no report of any patient involvement or harm as a result of this issue.The philips fse went to the customer¿s site to evaluate the ct system.The fse was able to confirm the customer¿s allegation and found that the breathing light assembly had failed.The fse replaced the breathing light assembly and a ghost to audio board cable to resolve the issue.The fse completed a functional check, which passed, and then returned the system back to the customer for clinical use.The system is operational and in clinical use.This event has been determined not to be a reportable event.
|
|