• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIO-DETEK INCORPORATED ONE STEP PEDIATRIC CPR ELECTRODE; ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIO-DETEK INCORPORATED ONE STEP PEDIATRIC CPR ELECTRODE; ELECTRODES Back to Search Results
Model Number 8900-000219-01
Medical Device Problem Code Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event or Problem Description
Complainant alleged that during a routine shift check by a clinician, the associated device displayed "defib pad short" message using these electrodes pads.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
The electrodes were returned for evaluation; the malfunction was duplicated during visual evaluation.The malfunction was attributed to a disconnected jumper wire (self test wire) on the electrode pad.This type of failure does not disable the electrode from delivering therapy when attached to a defibrillator.This is a malfunction which only impacts self-test of the electrode with the defibrillator.This type of malfunction does not pose any clinical impact and does not meet our guidelines for reportability.Analysis for reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONE STEP PEDIATRIC CPR ELECTRODE
Common Device Name
ELECTRODES
Manufacturer (Section D)
BIO-DETEK INCORPORATED
525 narragansett park drive
pawtucket RI 02861
MDR Report Key7437481
Report Number1218058-2018-00031
Device Sequence Number196023
Product Code MKJ
UDI-Device Identifier00847946023348
UDI-Public00847946023348
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/15/2019
Device Model Number8900-000219-01
Device Catalogue Number8900-000219-01
Device Lot Number2917
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2018
Initial Date Received by Manufacturer 03/28/2018
Supplement Date Received by Manufacturer03/28/2018
Initial Report FDA Received Date04/17/2018
Supplement Report FDA Received Date05/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-