Catalog Number PTAX4-14-170-4-16
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that when the ptax4-14-170-4-16 advance 14 lp low profile balloon catheter was removed from it's packaging, material which was described as a film on the surface of the balloon, appeared to be coming off it.The balloon catheter was not used for the procedure.A second ptax4-14-170-4-16 advance 14 lp low profile balloon catheter was used to successfully complete the procedure.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence nor were any additional procedures required.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: a review of the complaint history, device history record, documentation, drawing, instructions for use (ifu), manufacturing instructions, quality control, specifications, and a visual inspection and functional evaluation of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed the presence of a clear, stringy substance on the balloon.No defects were noted on the balloon catheter.The color of the balloon was transparent.The balloon dimensions were measured to be within quality specifications.The balloon was inflated to 16 atmospheres without any leakage or rupture occurring.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed the lot was released meeting all finished goods release criteria.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, the root cause was determined to be related to the device manufacturing process.Measures are being conducted to address this failure mode.We will continue to monitor for similar complaints.Per the quality engineering risk assessment, no further action is required.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 14 LP LOW PROFILE BALLOON CATHETER
Common Device Name
DQY CATHETER, PERCUTANEOUS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7437540
Report Number1820334-2018-00989
Device Sequence Number1007124
Product Code DQY
UDI-Device Identifier10827002503394
UDI-Public(01)10827002503394(17)201116(10)8379682
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K090822
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPTAX4-14-170-4-16
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 03/28/2018
Supplement Date Received by Manufacturer07/18/2018
Initial Report FDA Received Date04/17/2018
Supplement Report FDA Received Date07/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-