| Catalog Number |
140.655 |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Code |
Tissue Damage (2104)
|
| Date of Event |
02/23/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.
|
| |
|
Event or Problem Description
|
|
The clinician reported that 4 months after the dental implant was placed in ada site 19, the implant had not achieved osseointegration in type ii bone.Clinician reports mobility during torquing.The product will be forwarded to the manufacturer for investigation.There were no reported complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.
|
| |
|
Event or Problem Description
|
|
The clinician reported that 4 months after the dental implant was placed in ada site 19, the implant had not achieved osseointegration in type ii bone.Clinician reports mobility during torquing.The product will be forwarded to the manufacturer for investigation.There were no reported complications.(b)(4).
|
| |
|