| Model Number |
H7493926032220 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older.Device is a combination product.(b)(4).Device evaluated by mfr: stent delivery system was returned for analysis.A visual examination of the stent found damage to distal stent rows 1-7 with stent struts lifted and pulled in a distal direction over the distal markerband.The undamaged crimped stent outer diameter was measured and the result was within maximum crimped stent profile measurement the balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple hypotube kinks along the full length of the catheter.This type of damage is consistent with excessive force being applied to the delivery system.A visual and tactile examination of the outer and inner lumen and mid-shaft section found multiple inner/outer polymer extrusion kinks.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design and manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 27-march-2018.It was reported that crossing difficulties were encountered.A 2.25 x 32mm synergy ii drug-eluting stent was advanced but failed to cross the lesion.The procedure was completed with another of the same device.No patient complications were reported.However, returned device analysis revealed distal stent damaged.
|
| |
|
Search Alerts/Recalls
|