• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS HICKMAN 9 FR DUAL-LUMEN CV CATHETER, CUTDOWN TRAY WITH SURECUFF TISSUE INGROWTH; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS HICKMAN 9 FR DUAL-LUMEN CV CATHETER, CUTDOWN TRAY WITH SURECUFF TISSUE INGROWTH; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number 0600330
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer, at this time, for evaluation.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.Device not returned at this time.
 
Event or Problem Description
It was reported that the ¿patient was due iv flucloxacillin antibiotics made up into 20mls.When attempting to bolus the drug into the white port, liquid appeared to squirt out of the side of the actual line.On examination a small fracture approximately 2mm in length could be seen just below the clamp.¿ hickman catheter.No patient injury reported.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of fractured catheter was confirmed and the cause appears to be use related.The product returned for evaluation was one 9 fr dual lumen hickman catheter.The damage observed in the returned sample was characteristic of over-pressurization (burst) damage.This can occur through the use of syringes smaller than 10ml, by flushing against an occlusion, or due to excessive force applied during infusion procedures.The returned product sample was evaluated and a split was observed immediately distal of the clamping sleeve.This catheter damage was typical of a burst, and the characteristics observed which supported this type of failure included: 'c' shaped split.Tensile weakness at the fracture site (due to material ballooning prior to burst).Granular fracture surface texture (typical of tearing failure modes).An occlusion was observed distal to the burst site, and this may have contributed to the failure.Forceful flushing against an occlusion can cause this type of failure.An examination of the catheter structure revealed no potential damage/defect related to manufacture of the product.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Event or Problem Description
It was reported that the ¿patient was due iv flucloxacillin antibiotics made up into 20mls.When attempting to bolus the drug into the white port, liquid appeared to squirt out of the side of the actual line.On examination a small fracture approximately 2mm in length could be seen just below the clamp.¿ hickman catheter.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HICKMAN 9 FR DUAL-LUMEN CV CATHETER, CUTDOWN TRAY WITH SURECUFF TISSUE INGROWTH
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7438564
Report Number3006260740-2018-00728
Device Sequence Number13743321
Product Code LJS
UDI-Device Identifier00801741051753
UDI-Public(01)00801741051753
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K830233
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0600330
Device Catalogue Number0600330
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2018
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer05/14/2018
Initial Report FDA Received Date04/17/2018
Supplement Report FDA Received Date05/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-