Device used for treatment, not diagnosis.Event date: unknown when non-union began.Complainant part is not expected to be returned for manufacturer review/investigation.(b)(4) used for: hardware removal of a hind-foot nail took place on (b)(6) 2018 due to nonunion of subtalar joint.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device history records review not available for reporting at this time.Based on the information available, this complaint will be accounted for and monitored via post market surveillance activities.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: part: 04.008.016s; lot: 7792858; manufacturing location: selzach; supplier: (b)(4); release to warehouse date: november 03, 2014 non-sterile 04.008.016 / 7792858 was manufactured in us, monument.Manufacturing date: september 17.2014.Part #: 04.008.016, lot#: 7792858, 10mm ti hindfoot arthrodesis cann nail-ex 180mm/right ¿ qty: 6.No non-conformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Components parts reviewed: raw material part 21012 bp80 lot 7691972.Raw material provided by perryman company.Certified test report for titanium ingot received from perryman meet specification.Raw material receiving/putaway checklist meet specification.Inspection sheet for in-process acceptance sheet met inspection acceptance criteria.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|