• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE IOL; INTRAOCULAR LENS Back to Search Results
Model Number CQ2015A
Medical Device Problem Codes Improper or Incorrect Procedure or Method (2017); Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: collamer ultraviolet-absorbing posterior chamber single piece foldable intraocular lens.(b)(4).
 
Event or Problem Description
The reporter indicated that as the cq2015a intraocular lens, diopter +22.0 was being injected into the patient's eye, the trailing haptic portion of the lens tore off.Reportedly, this was due to technician error.The lens was successfully exchanged intraoperatively with an alternate lens.This occurred on (b)(6) 2018.
 
Additional Manufacturer Narrative
Device evaluation: the lens was returned in liquid, in a lens vial.Visual inspection found the optic torn, haptic torn, and the haptic is separated from the lens.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7438808
Report Number2023826-2018-00559
Device Sequence Number1591778
Product Code HQL
UDI-Device Identifier00841542102025
UDI-Public00841542102025
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
P990013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date07/31/2019
Device Model NumberCQ2015A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/22/2018
Supplement Date Received by Manufacturer05/04/2018
Initial Report FDA Received Date04/17/2018
Supplement Report FDA Received Date05/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age61 YR
-
-