| Catalog Number |
AR-1204.5L |
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
03/20/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.Device history record review revealed nothing relevant to this event.The device was requested for evaluation but was not returned therefore the complainant's event could not be verified.The cause of the event could not be determined from the information available and without device evaluation.Complainant's event typically caused by user mechanical damage to device such as hitting the device with another device, prying/leveraging or excessive bending forces being applied during use.The potential causes of this event are being communicated to the event reporter.If the device is returned and additional information is obtained, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the 4.5mm cannulated drill has a broken tip.No further details provided.Additional information has been requested.Follow-up investigation: the surgeon was performing a deep deltoid ligament reconstruction through bony tunnels with hamstring autograft procedure and as she was drilling into a space with existing hardware, she hit a screw.The tip of the drill broke-off and was left in bone, as a broken screw would be.Only the base of the drill will be returned for evaluation, the tip was left in the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that the 4.5mm cannulated drill has a broken tip.No further details provided.Additional information has been requested.Follow-up investigation: the surgeon was performing a deep deltoid ligament reconstruction through bony tunnels with hamstring autograft procedure and as she was drilling into a space with existing hardware, she hit a screw.The tip of the drill broke-off and was left in bone, as a broken screw would be.Only the base of the drill will be returned for evaluation, the tip was left in the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.Device history record review revealed nothing relevant to this event.Complaint confirmed.Visual examination revealed the distal tip of the drill broke off laterally down the shaft to just proximal to the 10mm mark.The od, material type, and hardness were all within specification.The broken fragment was not received for evaluation.Complainant's event typically caused by user mechanical damage to device such as hitting the device with another device as stated in the event description.The potential causes of this event are being communicated to the event reporter.If the device is returned and additional information is obtained, a follow-up report will be submitted.
|
| |
|