ARTHREX INC. BURR, ROUND, HL, 8 FLUTE, 4.0MM X 19CM; BUR, SURGICAL, GENERAL & PLASTIC SURGERY
|
Back to Search Results |
|
| Catalog Number |
AR-6400RBE |
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
03/30/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.Device history record review revealed nothing relevant to this event.The device was requested for evaluation but was not returned therefore the complainant's event could not be verified.The cause of the event could not be determined from the information available and without device evaluation.Complainant's event typically caused by user mechanical damage to device such as hitting the device with another device or excessive bending forces being applied during use.The potential causes of this event are being communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a hip arthroscopy procedure, the surgeon noticed that the 40 mm round burr was splitting and producing metal shavings in the joint.A second burr of the same lot was opened and used and again, the burr split and produced metal shavings.The surgeon was able to remove most of the shavings via suction and the case was completed.
|
| |
|