| Model Number |
T16X12 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
03/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product involved in the report has been returned and is being processed for evaluation.Upon completion of the investigation; a follow-up report will be filed.All information reasonably known as of 14apr2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the doctor performed and initial placement on a patient.The patient was in the right lateral recumbent position with no issues upon the insertion.The catheter was placed bottom to top to a depth of 15 cm and ropivacain was used to bolus.When the sheath was peeled away the healthcare professional felt like it broke leaving a piece behind.The doctor removed the catheter in an attempt to retrieve the broken piece, however, that was unsuccessful.An ultrasound was used to verify if there was a missing piece and the ultrasound confirmed the retained component inside the patient.The doctor felt that a secondary incision was necessary to locate the remaining piece.A 4¿ incision was made, but the healthcare professionals were unable to locate the piece.Rather than cause more trauma to the area, the team decided to abort and let the surgeon try to perform the rib fracture repair.The next day the patient underwent surgery to fix the rib fracture.The doctor performed the surgery, but could not locate or see the missing piece.The patient experienced a normal recovery despite not locating the missing piece.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
One used sample was returned.The t-peel sheath was received with the finger tabs separated from the lumen portion of the sheath.The entire portion of the sheath was not returned.Small strands of stretched sheath were attached to each of the finger tabs but it did not appear that when the customer had pulled on each of the tabs that it did not separate the sheath longitudinally into two even pieces.The investigation summary concluded that the sheath was broken.The tabs had been pulled but separated form the lumen portion of the sheath and had not separated it into two sections as is designed to do.No root cause was identified.All information reasonably known as of 25-jun-2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a halyard health product is defective or caused serious injury.
|
| |
|
Search Alerts/Recalls
|