• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number EGIA30AV
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/07/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during the first firing of the device in laparoscopic nephrectomy procedure, they connected reload to the handle.The surgeon clamped the renal vein and pushed the green button and tried to squeeze the handle ,however could not squeeze it.They replaced it with a new reload, and the surgeon clamped the renal vein, however the same event occurred.The sales rep noted some staples of the proximal end were pre-fired.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of two devices.Visual inspection noted that the first reload was pre-fired and engaged in interlock with the jaws open.The second reload was received with a full complement of staples.Functional testing of the reloads found no abnormalities.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the pre-fire condition with interlock engagement may occur if the instrument firing handle had been partially compressed and released after pressing the green firing button.In this situation, the safety interlock feature will engage and prevent the reload from firing a second time by ceasing the placement of staples and tissue transection and prevent patient harm.The root cause of the observed pre-fired reload was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened, and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during the first firing of the device in laparoscopic nephrectomy procedure, they connected reload to the handle.The surgeon clamped the renal vein and pushed the green button and tried to squeeze the handle ,however could not squeeze it.They replaced it with a new reload, and the surgeon clamped the renal vein, however the same event occurred.They then replaced the devices with a new handle and a new cartridge, and the firing was completed with no problem.The sales rep noted some staples of the proximal end were pre-fired.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott road
north haven CT 06473
MDR Report Key7439927
Report Number1219930-2018-02075
Device Sequence Number13710861
Product Code GDW
UDI-Device Identifier20884521186092
UDI-Public20884521186092
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Model NumberEGIA30AV
Device Catalogue NumberEGIA30AV
Device Lot NumberN7G0329KX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2018
Initial Date Received by Manufacturer 03/07/2018
Supplement Date Received by Manufacturer04/27/2018
Initial Report FDA Received Date04/18/2018
Supplement Report FDA Received Date05/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-