• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND ACCOLADE; IMPLANTABLE PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND ACCOLADE; IMPLANTABLE PULSE GENERATOR Back to Search Results
Model Number U128
Device Problems Failure to Capture (1081); High impedance (1291); Over-Sensing (1438); Ambient Noise Problem (2877)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/15/2018
Event Type  malfunction  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy pacemaker (crt-p) triggered a lead safety switch on the right ventricular (rv) lead.The switch occurred due to high out of range rv pace impedances greater than 2000 ohms.Prior to the switch, there was also an episode of oversensed noise.Troubleshooting was performed and in bipolar, the out of range impedances were observed and there was also loss of capture.In unipolar, all measurements were stable.Boston scientific technical services discussed troubleshooting.At this time, the products remain in service.The system was reprogrammed to bipolar and the patient is being monitored.No adverse patient effects were reported.
 
Event Description
Additional information was received which indicated another review was performed of this device data by boston scientific technical services.The minute ventilation signal appeared sensed in an episode, but it did not impact therapy.At this time, there have been no adverse patient effects reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE
Type of Device
IMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7440037
MDR Text Key105886880
Report Number2124215-2018-07141
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559402
UDI-Public00802526559402
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 09/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/29/2018
Device Model NumberU128
Other Device ID NumberVALITUDE X4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received09/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/29/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age77 YR
-
-