| Model Number |
RELION TMA |
| Medical Device Problem Code |
Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned meter evaluated with no defect found.Returned test strips evaluated with defect found; black chemistry observed.Qc investigation on 4/12/2018 resulted in defect found: no chemistry observed on returned test strips.After qc investigation the event was determined to be reportable.Final root cause: other root cause: washed strips.Test strip udi# (b)(4).Manufacturer contacted customer (several attempts) in a follow-up call in order to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer complaint of e-0 error; test strip error.E-0 error occurred when the blood sample is absorbed.Customer feels well and observed no symptoms.Medical intervention is not reported at the time of the call on (b)(6) 2018 as a result of the meter's readings.Blood test performed fasting during call on (b)(6) 2018 produced result of e-0 upon sample of blood being applied to strip.Storage location of product is undisclosed.Test strip lot manufacturer's expiration date is 07/22/2019 and open vial date is one week ago.
|
| |
|
Search Alerts/Recalls
|