| Model Number |
R SERIES |
| Medical Device Problem Code |
Failure to Discharge (1169)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device failed to discharge using external paddles ((b)(4)).Complainant indicated that there was no patient involvement in the reported malfunction.Please reference medwatch reports 1220908-2018-01060 and 1220908-2018-01063 for similar events reported from the same customer.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned to zoll medical corporation; the device's external paddles were removed and returned.The malfunction was duplicated and attributed to a faulty connector on the external paddles.Analysis for reports of this type has not identified an increase in trend.
|
| |
|