Device was used for treatment, not diagnosis.Patient age, date of birth, gender, and weight not available for reporting.Date when non-union began is unknown.Device is not expected to return.Device history record review: manufacturing location: (b)(4), manufacturing date: 31-jul-2015, expiration date: 30-jun-2025.Part #: 04.037.028s, lot#: 9864913 (sterile) - 10 mm/125 deg ti cann tfna 380 mm/right- sterile.Quantity 6.Inspection sheet for in-process/inspect dimensional/final met inspection acceptance criteria.Component parts reviewed: the 04.037.912.2 - lock prong 125 degree, tfna bp-55 lot - 9529795.The 04.037.912.4 - wave spring, shim ended bp-55 lot - 7722133.The 4.037.912.3 - tfna lock drive bp-58 lot - 9841103.The 21127 - raw material lot bp-80 lot - 7813438.Raw material received from (b)(4).Certified test report received from (b)(4), certificate of test for titanium received from (b)(4) and raw material receiving/putaway checklist meet specification.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.(b)(4), ethicon (abq), ¿sterility documentation was reviewed and determined to be conforming.¿ the investigation could not be completed; no conclusion could be drawn, as no product was received.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|