• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER POLYFLUX 21L; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER POLYFLUX 21L; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 104176
Device Problems Fluid/Blood Leak (1250); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that there was backpressure in a polyflux 21l dialyzer, which led to blood leakage during treatment.A non-baxter dialysis machine was being used at the time of the event.Upon follow up, the customer stated that the lines had not been tightened properly, leading to the leak.The reporter indicated that the patient may have received prophylactic antibiotics.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information: event.Event - baxter renal therapy specialist visited the clinic and reviewed proper procedure with the staff.After reviewing the procedure, the customer indicated that some new staff members had not been tightening the lines properly.The customer has now introduced a second check for the tightness of connections during preparation.The cause of this event could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYFLUX 21L
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER
hechingen
MDR Report Key7441039
MDR Text Key106336734
Report Number9611369-2018-00040
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
PMA/PMN Number
K043342
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number104176
Device Lot Number7-6041-H-01
Was Device Available for Evaluation? No
Date Manufacturer Received04/22/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
BRAUN HD MACHINE
-
-