| Model Number |
5640 |
| Medical Device Problem Code |
Detachment Of Device Component (1104)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
05/18/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being based on the visual investigation and condition of the returned sample.A manufacturing record review was completed and zero nonconformances were found.The returned product evaluation confirmed the damage to the distal tip section.There was a longitudinal tear in the distal tip but everything remained intact.Also, the mid-shaft of the catheter was kinked.The damage is consistent with over torquing against resistance.
|
| |
|
Event or Problem Description
|
|
Doctor was unsuccessful in resolving this cto and will be attempting again in two months time.Could not progress through a retrograde approach and then proceeded with an antegrade approach.Could not penetrate the proximal cap.The turnpike gold was bent at a 45 degree angle with a wire stuck in it.The wire and turnpike had to be removed through the sheath together.Dr.Was attempting to navigate through the proximal cap of a mid.Circ.Lesion, 9 different wires were used, and 5 various turnpikes were also used.The tips of the lp and spiral catheters were deformed and had difficulty removing the wire.This report is associated to medwatch 2134812-2018-00022.
|
| |
|