Model Number MMT-723LNAH
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/06/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The insulin pump was received with missing segment/partial display, problem isolated to lcd board.The insulin pump had minor scratched lcd window, cracked case, cracked lcd window, broken battery tube threads, cracked reservoir tube lip, cracked case at display window corners, cracked case at reservoir tube window corners, stained address/serial number label and missing end cap sticker.The insulin pump was received without original battery cap.Unable to verify damage to the original battery cap.
 
Event or Problem Description
Customer reported via phone call for cosmetic damage.Blood glucose was unknown.The customer also reported that the insulin pump was taped together.Customer reported that her battery cap was broken and there are cracks above the screen.The insulin pump was returned for analysis.
 
Additional Manufacturer Narrative
The aware date provided with the initial report was incorrect.The correct information has been provided with this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7441175
Report Number3004209178-2018-65957
Device Sequence Number16434294
Product Code OYC
UDI-Device Identifier00643169478572
UDI-Public(01)00643169478572
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 04/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723LNAH
Device Catalogue NumberMMT-723LNAH
Device Lot NumberA4723LNAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2018
Initial Date Received by Manufacturer 12/06/2017
Supplement Date Received by Manufacturer04/05/2018
Initial Report FDA Received Date04/18/2018
Supplement Report FDA Received Date04/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age26 YR
Patient Weight150
-
-