| Model Number |
7266V |
| Medical Device Problem Code |
Hole In Material (1293)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/01/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being based on the visual investigation and condition of the returned sample.A manufacturing record review was completed and zero nonconformances were found.The returned product evaluation confirmed the puncture/hole in the sheath.The sheath was kinked at the section where the hole was present.Based on the event details the most likely root cause for the failure is the mandrel punctured the sheath as the doctor was trying to advance the sheath.
|
| |
|
Event or Problem Description
|
|
Dr.Was using mik for access in a varilase case.When he tried to advance the 035"" j-tip wire through the mik sheath, he felt resistance.Dr.Than removed the j-tip wire, and used the mik mandrel to advance and removed the mik sheath.The tip of the sheath showed a puncture at the distal tip.Mik sheath removed and placed in biohazard bag with packaging.No patient impact or injury reported.
|
| |
|