• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS ARTOURA BREAST TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS ARTOURA BREAST TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE Back to Search Results
Catalog Number SMXP140RUH
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 01/03/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The mentor failure analysis lab has received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A device history record review is in progress.Once completed, a supplemental report will be submitted.Concomitant medical products: left unknown mentor tissue expander.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) female patient underwent a primary breast reconstruction with mentor artoura 850cc tissue expanders which the right side deflated after implantation.As a result, patient had the device removed and replaced with another smooth artoura uhp 850cc tissue expander, serial number (b)(4) on (b)(6) 2018.Beside the surgery, no other symptoms observed.
 
Additional Manufacturer Narrative
Device evaluation summary upon receipt by mentor, the device returned with clear fluid.No foreign material was observed within the device or on the shell surface.Product evaluation team discovered a rent measuring approximately < 0.1 cm on the posterior aspect.Because the authorization for return and examination of medical device form was not signed and/or returned to mentor, product evaluation team was precluded from further evaluating the device.No other anomalies were discovered.The complaint was confirmed since a rupture was found on the device.However, at this point it is not possible to determine the root cause of such damage.All the implants are 100% inspected before leaving the facility.There is no evidence that the issue is related with manufacturing.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On 6/6/2018, the device history record (dhr) was reviewed, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTOURA BREAST TISSUE EXPANDER
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key7441732
Report Number1645337-2018-02329
Device Sequence Number13710725
Product Code LCJ
UDI-Device Identifier00081317024415
UDI-Public00081317024415
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/15/2021
Device Catalogue NumberSMXP140RUH
Device Lot Number7480048
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2018
Initial Date Received by Manufacturer 03/22/2018
Supplement Date Received by Manufacturer03/22/2018
03/22/2018
Initial Report FDA Received Date04/18/2018
Supplement Report FDA Received Date05/10/2018
06/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age40 YR
-
-