| Catalog Number |
100/875/070 |
| Medical Device Problem Codes |
Suction Problem (2170); Defective Component (2292)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the suction link of a portex® blue line ultra® suctionaid® tracheostomy tube would not work.The fault was noticed immediately upon use with the patient.No injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One portex® blue line ultra® suctionaid® tracheostomy tube was returned for analysis in used condition.The connector was noted to be occluded / covered by plastic upon visual exam.It was noted that the plastic covering the connector was from the same material as the connector.Relevant documents were reviewed and deemed adequate (mp trachy (b)(4) blus tracheostomy tube-bond connector to suction line tube; qp trachy (b)(4) trachy blue line assembly and packaging).An audit of the production floor was conducted by the quality engineer was performed.Line clearance records, training records and production personnel are following the procedure mp trachy (b)(4) blus tracheostomy tube-bond connector to suction line tube were reviewed with no discrepancies.Flow testing of (b)(4) units were audited detecting occlusions; confirming complaint.Based on the evidence, the possible root cause is from supplier from the following findings: the connector was received occluded from the supplier; lack of detection by production personnel.
|
| |
|