| Model Number |
MMT-723NAL |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump alarmed a33 compromised force sensor system.The customer¿s blood glucose level was unknown at the time of the incident.Customer stated that the alarm occurred when tried to fill the reservoir and place in pump prime rewind.The customer stated that the drive support cap was protruded/ loose/sticking out.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump passed the displacement test but compromised force sensor system alarm during the basic occlusion test due to loose or protruded drive support disk.Unable to perform the occlusion, prime or compromised force sensor system and excessive no delivery test due to compromised force sensor system alarm.
|
| |
|