• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC THERMOCOOL® SF BI-DIRECTIONAL CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC THERMOCOOL® SF BI-DIRECTIONAL CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Catalog Number BDI35DFRT
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Cardiac Tamponade (2226)
Date of Event 03/21/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product was discarded, therefore no product failure analysis can be conducted and device malfunction cannot be confirmed.Device history record (dhr) review cannot be conducted because no lot number was provided by the customer.Biosense webster manufacturer's reference number (b)(4) has two complaints that are related to the same incident.Manufacturer's ref.No: (b)(4).
 
Event or Problem Description
It was reported that a patient underwent an accessory pathway ablation procedure with an ez steer¿ bi-directional electrophysiology catheter and thermocool® sf bi-directional catheter and suffered a cardiac tamponade.During the accessory pathway procedure, a cardiac tamponade was detected.There is no information regarding interventions, extended hospitalization, patient outcome, factors that may have contributed to the adverse event, or the physician¿s causality opinion.There were no errors reported on any biosense webster, inc.Products or equipment during the procedure.There is no information regarding transseptal puncture, sheath, generator parameters, generator settings, power titration, overall ablation time at the site of injury, last ablation cycle time at the site of injury, irrigated catheter flow rate, or anticoagulation during the procedure.This was not a carto system case, as the catheters were non-navigational.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SF BI-DIRECTIONAL CATHETER
Common Device Name
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
33 technology drive
irvine CA 92618
MDR Report Key7442337
Report Number2029046-2018-01447
Device Sequence Number19106176
Product Code LPB
UDI-Device Identifier10846835002812
UDI-Public10846835002812
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P030031/S025
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberBDI35DFRT
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/22/2018
Initial Report FDA Received Date04/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
-
-