• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT HEMISPHERICAL SOLID BACK SHELL; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT HEMISPHERICAL SOLID BACK SHELL; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 500-11-48D
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Not returned.
 
Event or Problem Description
Nurse informed that the inside of the shell has a scratch.Customer used another shell.
 
Additional Manufacturer Narrative
Corrected data: - expiration date.Additional information: - gtin; product available to stryker.An event regarding appearance issue involving a trident shell was reported.The event was not confirmed.Method & results: device evaluation and results: visual inspection -- a visual inspection was performed on the returned device noted that no scratches were present on the inside or outside of the device.Dimensional inspection: not performed as reported event is not related to the dimensional integrity of the device.Functional inspection: not performed as reported event is not related to device function.Material analysis: not performed as reported event is not related to material integrity.Medical records received and evaluation: a medical review was not performed because no medical information was provided.Device history review: review indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: review indicated there have been no other similar events for the reported lot.Conclusions: the event could not be confirmed.There were no scratches present on the returned device.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Nurse informed that the inside of the shell has a scratch.Customer used another shell.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT HEMISPHERICAL SOLID BACK SHELL
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7442547
Report Number0002249697-2018-01165
Device Sequence Number1769400
Product Code LPH
UDI-Device Identifier04546540431844
UDI-Public04546540431844
Combination Product (Y/N)N
Initial Reporter CountryFI
PMA/510(K) Number
K143085
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 06/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2023
Device Catalogue Number500-11-48D
Device Lot Number62811401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2018
Initial Date Received by Manufacturer 03/23/2018
Supplement Date Received by Manufacturer05/16/2018
Initial Report FDA Received Date04/19/2018
Supplement Report FDA Received Date06/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-