An internal biomérieux investigation was performed with results as follows: a k.Pneumoniae isolate was submitted for esbl evaluation because the vitek® 2 ast-gn67 card gave negative esbl test results on multiple runs.When the isolate was tested on microscan walkaway, an esbl positive result was obtained.Identification was confirmed, and testing included both the customer lot and a random lot of ast-gn67 cards, run in duplicate.The clsi esbl disk test, which is the reference method, was also performed.On all four (4) card runs, negative esbl test results were obtained.The clsi esbl disk test was also negative.Therefore, the customer's negative esbl test results were duplicated in-house and are in agreement with the reference method.Cards are performing as expected.
|