| Catalog Number |
0006428 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The full number for the product in question is (b)(4) (1/27/2003).Immucor technical support used a remote electronic connection method on (b)(6) 2018 to assess the instrument test well images in question, of which some appeared as visually positive (homozygous m antigen expression test wells) and some appeared as visually negative (heterozygous m antigen expression test wells).The immucor laboratory tested retention product on (b)(6) 2018, which performed as expected.Complaints of positive reactions being interpreted as negative by the galileo echo camera algorithm are being corrected under design control project 13-005.The immucor technical communication cc-09-042-02 is being used by the lab to visually confirm all negative assay events reported by the echo instrument.The internal immucor tracking number for this report is (b)(4).
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, a customer site reported an unexpectedly negative antibody screen when using capture-r ready indicator red cells by galileo echo method.
|
| |
|
Search Alerts/Recalls
|