| Model Number |
86112 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
Extubate (2402)
|
| Date of Event |
04/05/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during use, the device's cuff did not inflate when an air was pushed.The customer indicated that the patient had to be re-cannulated with a new tube.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one sample was received for analysis and the reported issue could not be confirmed.A visual inspection was performed and it was observed all the components are assembled properly at the tube.An inflation/deflation test was performed using a syringe 25cc of air was applied to the cuff and it was observed the cuff held the air, the sample was left inflated 2 hours according to process instruction and after this time no deflation was observed.The sample was submerged into a container filled with water and no leaks were observed.If information is provided in the future, a supplemental report will be issued.
|
| |
|