• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEODENT - JJGC S.A. CM DRIVE IMPL TI 5.0X16; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEODENT - JJGC S.A. CM DRIVE IMPL TI 5.0X16; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 109.687
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Tissue Damage (2104)
Date of Event 02/07/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Exemption number: e2015015.
 
Event or Problem Description
The clinician reported that almost 6 months after the dental implant and abutment were placed in ada site 3, the implant had not achieved osseointegration in type i bone clinician reports bone loss.The product will be forwarded to the manufacturer for investigation.There were no reported complications.
 
Additional Manufacturer Narrative
Exemption number: e2015015.Considering the surgical methodology, it was seen that the dentist has used less drills than recommended to perform the implant installation.
 
Event or Problem Description
The clinician reported that almost 6 months after the dental implant and abutment were placed in ada site 3, the implant had not achieved osseointegration in type i bone clinician reports bone loss.The product will be forwarded to the manufacturer for investigation.There were no reported complications.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CM DRIVE IMPL TI 5.0X16
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
NEODENT - JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key7443667
Report Number3008261720-2018-01823
Device Sequence Number19090181
Product Code DZE
UDI-Device Identifier07898237569253
UDI-Public07898237569253
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K123022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Dental Assistant
Type of Report Initial,Followup
Report Date (Section B) 04/19/2018
Report Date (Section F) 04/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Dentist
Device Expiration Date01/30/2020
Device Catalogue Number109.687
Device Lot Number800173200I
Other Device ID Number07898237569253
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2018
Was the Report Sent to FDA? No
Distributor Facility Aware Date07/02/2018
Device Age2 MO
Event Location Other
Date Report to Manufacturer04/11/2018
Initial Date Received by Manufacturer 04/11/2018
Supplement Date Received by Manufacturer06/12/2018
Initial Report FDA Received Date04/19/2018
Supplement Report FDA Received Date07/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ABUTMENT REMOVAL; ABUTMENT REMOVAL; ABUTMENT REMOVAL
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
-
-