Blank fields on this form indicate the information is unknown, or unavailable.Additional information: patient's age: 37 years; date of birth: 10/13/1980.Date of event: (b)(6) 2017.The customer confirmed that there were no difficulties experienced during the initial placement of the stent, and that the usual products and equipment used during this type of procedure was employed in this case as well.The date of implantation was reported to be (b)(6) 2017, and the stent was removed on (b)(6) 2017.Implant date: (b)(6) 2017; explant date: (b)(6) 2017.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Investigation evaluation: a review of the complaint history, device history record, instructions for use (ifu), manufacturing instructions, and quality control of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there was one other reported complaint for this lot number.The device is shipped with instruction for use (ifu) which notes: when choosing a stent length, ensure that the length chosen will position the mushroom tips above and below the biliary lesion to be stented.Position the stent across the obstructing lesion, with biliary lesion centered between mushroom tips.It is not known if there was a ¿delay in stent introduction at least three to seven days after initial insertion of the internal/external catheter into the duodenum as per the ifu.Based on the information provided, no product returned and the results of our investigation, the possible root cause of this complaint is related to the patient anatomy, (possibly related to the pre-existing patient condition.The patient had pre-existing cholangitis stricture of the cbd) however a root cause could not be determined with certainty.We will continue to monitor for similar complaints and notify appropriate personnel.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|