ENDOLOGIX INC. AFX2; AFX2 BIFURCATED STENT GRAFT Back to Search Results
Model Number BEA22-60/I13-40
Medical Device Problem Codes Partial Blockage (1065); Off-Label Use (1494)
Health Effect - Clinical Code Occlusion (1984)
Date of Event 07/26/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It is anticipated that the device remains in the patient.If additional information pertinent to the event is obtained, a follow-up report will be submitted.
 
Event or Problem Description
On (b)(6) 2016, the patient was initially implanted with an afx2 bifurcated stent graft.On (b)(6) 2017, a secondary procedure was performed for an unknown reason.During the secondary procedure a afx vela infrarenal was implanted.No additional event or patient information is available.
 
Event or Problem Description
Subsequent to the initial mrd, additional information was received.The implanting physician for the primary procedure performed on (b)(6) 2016 was dr.(b)(6).Reportedly, the patient developed a narrowing in the aorta above the afx2 graft.The secondary procedure on (b)(6) 2017, was performed by dr.(b)(6) to extend up the renals.Both procedures were performed at (b)(6) hospital.No additional event or patient information is available.
 
Additional Manufacturer Narrative
Based on the information received, clinical assessment determined the narrowing above the stent graft complaint is refuted rather there was a complete occlusion of the aorto-illiac artery.Procedure related harms for this complaint could not be determined based on the information shared with endologix to review.The final patient status was reported to be in stable post the secondary endovascular procedure.A definitive root cause was not determined; however, the most likely cause of the occlusion was pre-existing occlusive aorto-illiac disease which was observed in implant angiogram.In addition the off-label use of stent placed in non-aneurysmal aorta.The device remains in the patient and will not be returned for evaluation.The review of manufacturing lot confirmed all devices met specifications prior to release.This complaint is not capa eligible at this time.These types of events will be monitored and trended as part of the quality system.Initial reporter name: correction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
AFX2 BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX INC.
2 musick
irvine CA 92618
MDR Report Key7444341
Report Number2031527-2018-00319
Device Sequence Number11075812
Product Code MIH
UDI-Device Identifier00818009015068
UDI-Public(01)00818009015068(17)170329
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date03/29/2017
Device Model NumberBEA22-60/I13-40
Device Lot Number1356049021
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/25/2017
Supplement Date Received by Manufacturer06/22/2018
Initial Report FDA Received Date04/19/2018
Supplement Report FDA Received Date07/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
-
-