| Model Number |
CRPLUS |
| Medical Device Problem Code |
Failure to Analyze Signal (1539)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
02/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio-control determined that the cause of the reported issue was process residue on capacitor c156, leg 3, of the analog pcb assembly that keeps the ecg preamp from recognizing a shockable rhythm.The customer was provided with a replacement device.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device displayed a service indicator.There was no patient use associated with the reported event.Upon evaluation of the customer¿s device, physio-control found that the it did not recognize a shockable rhythm.As a result, defibrillation would not be possible if it were necessary.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device displayed a service indicator.There was no patient use associated with the reported event.Upon evaluation of the customer¿s device, physio-control found that the it did not recognize a shockable rhythm.As a result, defibrillation would not be possible if it were necessary.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d4 of the initial medwatch report indicates: gtin/device identifier (di) number ¿ blank.Section d4 of the initial medwatch report should indicate: gtin/device identifier (di) number ¿ (b)(4).Section d4 of the initial medwatch report indicates: model # ¿ 15.Section d4 of the initial medwatch report should indicate: model # ¿ crplus.
|
| |
|