• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NATUS MEDICAL INC. DBA INTEGRA NEUROSCIENCES (CAMINO) OLM INTRACRANIAL PRESSURE MONITORING KIT; INTRACRANIAL PRESSURE MONITORING DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NATUS MEDICAL INC. DBA INTEGRA NEUROSCIENCES (CAMINO) OLM INTRACRANIAL PRESSURE MONITORING KIT; INTRACRANIAL PRESSURE MONITORING DEVICE Back to Search Results
Model Number 110-4B
Medical Device Problem Codes Unable to Obtain Readings (1516); Device Issue (2379)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Event or Problem Description
The following was reported by a foreign user facility: "the hospital informed the catheter stopped the measurement during the use" additional information has been requested.
 
Additional Manufacturer Narrative
Additional information: no patient injury.Four attempts were made to obtain additional information on this incident, however there has been no response: additional information requested: 04april2018-no response.Additional information requested: 19april2018-no response.Additional information requested: 02may2018-no response.Additional information requested 15may2018- no response.Natus completed their investigation on (b)(6) 2018.The results included the reference: device history record review.Review of past capas/ capa determination.Review of complaints history/trend analysis.Risk management review.Results: the device history record for the catheter was reviewed, there were no anomalies observed during the manufacturing, packaging or inspection of the device or accessories while in process.Capa complaint log 201 6, 2017 reviewed on 07-apr-2018 specific for the complaint issue under investigation indicates that there no capa investigation related to this catheter complaint.No capa has been issued in 2018.There were twelve (12) confirmed complaint during apr-2016 through mar-2018 under complaint code "integimproper function".The numbers of catheter sold during this time, (81036) shows an incidence rate of 0.015%.The failure mode resulted in a "delay in patient treatment" (ids-b-bt-l-ooi ).A review of the camino risk management file has identified the failure mode of "does not measure intracranial pressure accurately" (ids-b-bt-l-ooi ), this is based on the complaint description.It can be concluded that the impact of the failure of the 110-4bt catheter resulted in a "delay in patient treatment" (ids-b-bt-l-ooi ).: the reported customer complaint could not be replicated.Root cause: based on the results of this investigation, the root cause of the reported customer complaint could not be determined at this time.No corrective action required based on the evaluation above.Trends will be monitored for this or similar complaints.Multiple attempes were made to obtain additional information from the customer, however there has been no response.This complaint is now deemed closed, should additional information be received, the complaint may be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
(CAMINO) OLM INTRACRANIAL PRESSURE MONITORING KIT
Common Device Name
INTRACRANIAL PRESSURE MONITORING DEVICE
Manufacturer (Section D)
NATUS MEDICAL INC. DBA INTEGRA NEUROSCIENCES
5955 pacific center blvd.
san diego CA 92121
MDR Report Key7444531
Report Number2023988-2018-00014
Device Sequence Number1963738
Product Code GWM
UDI-Device Identifier10381780023623
UDI-Public10381780023623
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K102875
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 04/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number110-4B
Device Catalogue Number110-4B
Device Lot Number266190-B09
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/21/2018
Supplement Date Received by Manufacturer03/21/2018
Initial Report FDA Received Date04/19/2018
Supplement Report FDA Received Date05/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-